The Digital Fume Hoods VAV kit

Simplifying compliance to laboratory fume hood regulations and good operating practice

Download Brochure

There are multiple regulations governing safety devices during their entire life cycle

 

An employer must protect operators while they carry out their duties.

Several regulations and standards such as ASHRAE 110 and OSHA 29 CFR 1910.1450, govern device life cycle and use. Legal provisions provide for sanctions in case of non-compliance.

The Eurotherm Digital Fume Hood VAV kit aims to:

  • Meet fume hood technical operating requirements
  • Provide information relevant to the safety of operators
  • Ease maintenance practice and periodic testing

Reliance on accredited companies to obtain documented evidence that you have done your best to protect operators and ensure laboratory safety

Once installed on new or existing fume hoods, the Eurotherm Digital Fume Hood VAV kit, facilitates achieving accredited test certification¹ while helping maintain an efficient operational performance level.

“Accreditation ensures that certification, inspection, verification bodies, testing, and calibration laboratories fulfill all the requirements of the standards to undertake conformity assessment activities.” ²

¹ In many sectors certifications are issued by an accredited third-party body in accordance with the standards ISO/IEC 17065, ISO/IEC 17020, ISO/IEC 17021-1, ISO/IEC 17024, and ISO/IEC 17025 whose independence with respect to the certified object – product, service, organization or person – is verified and attested by an Accreditation Body

² Credit: https://www.accredia.it/en/accreditation/

Laboratory Fume Hoods Life Cycle

Laboratory safety assessment. An external, independent agency should run periodic assessments. It reviews equipment, lab and personal/collective protective equipment, emergency kits, ventilation, cleanliness, exit paths, chemical management, waste handling, signage, policies, training, and any other aspects related to safety, including the Laboratory Safety Standard Operating Procedures.

Laboratory energy assessment. The audit usually begins with a review of historical and current utility data and a benchmarking of the building’s energy use against similar facilities. The final report should contain a rating scorecard against the benchmark.

Advisory Subject Matter Experts can perform a Cold Eyes review to assess the current status of laboratory fume hoods in use. A non-binding, confidential report is released after the survey.

Service Support

  • Spare parts management
  • Calibrations
  • Compliance
  • Alarm & event remote management
  • Service desk
  • Service level agreement
  • Exchange web portal
  • Cloud hosting

Periodic tests

During the life cycle of a fume hood, businesses are required to demonstrate that fume hood performance complies to latest ANSI/ASHRAE-110 and EN 14175 guidelines, these include:

  • As Manufactured (AM) test, also known as Type test
  • As Installed (AI) test, also known as Commissioning test
  • As Used (AU) test, also known as Periodic test

The estimated average time to complete the AI or the AU testing is around 2 to 4 working hours per hood, dependent upon laboratory/system conditions.

What is ASHRAE and ANSI/ASHRAE 110-2016

Chemical fume hoods are critical to laboratory safety; ineffective or poorly configured fume hoods can be harmful to personnel working with potentially dangerous chemicals.

ASHRAE (American Society of Heating, Refrigerating and Air-Conditioning Engineers) is dedicated “to serve humanity by advancing the arts and sciences of heating, ventilation, air conditioning, refrigeration and their allied fields”. This mission statement is enacted by developing standards and guidelines to guide HVAC industry and relevant applications, including laboratory fume hoods.

First published in 1985 and revised in 1995, ASHRAE Standard 110 provides a method of testing the performance of laboratory fume hoods. This method of testing applies to conventional, bypass, auxiliary air, and variable-air-volume (VAV) laboratory fume hoods.

OSHA 3404-11R Laboratory Safety Guidance and 29 CFR 1910.1450 Standard

US Occupational Safety and Health Administration (OSHA) designed laboratory safety guidance 3404-11R which “is intended for supervisors, principal investigators and managers who have the primary responsibility for maintaining laboratories under their supervision as safe, healthy places to work and for ensuring that applicable health, safety and environmental regulations are followed.”
The guidance contains recommendations as well as descriptions of mandatory safety and health standards.

OSHA’s Occupational Exposure to Hazardous Chemicals in Laboratories standard, 29 CFR 1910.1450, requires that fume hoods be maintained and function properly when used, 29 CFR 1910.1450(e)(3)(iii).